Double-stranded DNA concentration in 96-well plates with PicoGreen (FLUOstar Omega plate reader)
1 Purpose and scope
This protocol measures the concentration of double-stranded DNA in many samples at once (DNA extracts, PCR products, libraries in 96-well format) with Quant-iT PicoGreen and the FLUOstar Omega plate reader. For single samples use EnGeL-MOLB-QUBIT; for purity use EnGeL-MOLB-NANODROP.
2 Principle
PicoGreen fluoresces strongly only when bound to double-stranded DNA. The fluorescence of each well is compared with a standard curve made from a dilution series of lambda DNA, so the result is specific to dsDNA. The lab uses half the volumes of the manufacturer’s assay (100 µL per well) with the same high-range standards.
3 Before you start
| Item | Details |
|---|---|
| Hands-on time | 30 min for a 96-well plate |
| Total time | 40 min |
| Sample volume | 2 µL per well |
| Prepare in advance | 1X TE and PicoGreen working reagent, fresh on the day, in the dark |
| Key safety points | See Safety and waste |
4 Materials
4.1 Equipment
- Fluorescence plate reader FLUOstar Omega (BMG Labtech) with protocol “PicoGREEN_LC_100” (excitation filter 485-12, emission filter 520)
- Pipettes (single and multichannel) and filter tips; vortex mixer; plate shaker (optional)
4.2 Consumables
- Black flat-bottom 96-well plates (Greiner 96 F-bottom, as set in the reader protocol)
- Plastic tubes only (PicoGreen binds to glass)
4.3 Reagents and kits
- Nuclease-free water
- Quant-iT PicoGreen dsDNA reagent and 20X TE buffer (Invitrogen) [TO CONFIRM: catalogue number]
- Lambda DNA standard (100 µg/mL, supplied with the PicoGreen kit)
5 Reagent preparation
- 1X TE: dilute 20X TE 1:20 in nuclease-free water (e.g. 2.5 mL 20X TE + 47.5 mL water for 50 mL).
- Lambda working stock (2 ng/µL = 2 µg/mL): dilute the 100 µg/mL lambda DNA 50-fold in 1X TE, e.g. 10 µL Lambda DNA + 490 µL 1X TE for 500 µL (the earlier lab version, 9 µL + 441 µL, gives 450 µL at the same concentration).
- PicoGreen working reagent: dilute 1:200 in 1X TE, fresh on the day, in the dark. Per 96-well plate, prepare 5 mL: 25 µL PicoGreen + 4 975 µL 1X TE (50 µL per well).
Standard curve (in duplicate)
PicoGreen standards.
| Standard | Lambda working stock (µL) | 1X TE (µL) | Working reagent (µL) | Concentration (ng/mL) |
|---|---|---|---|---|
| S1 | 50 | 0 | 50 | 1000 |
| S2 | 5 | 45 | 50 | 100 |
| S3 | 0.5 | 49.5 | 50 | 10 |
| Blank | 0 | 50 | 50 | 0 |
6 Procedure
6.1 Before starting
- Use filter tips at all steps; plastic tubes only.
- Keep the PicoGreen reagent and working reagent protected from light.
6.2 Plate and reading
- Pipette each standard into the black plate in duplicate (table in Reagent preparation).
- Samples: 2 µL sample + 48 µL 1X TE per well + 50 µL working reagent.
- Mix by pipetting or a short plate shake.
- Incubate for 2 min at room temperature, protected from light.
- Plate reader: Manage Protocol → “PicoGREEN_LC_100”; Start measurement → Gain Adjustment; then Start measurement. The reader protocol is a fluorescence-intensity endpoint read (Greiner 96 F-bottom plate; excitation filter 485-12, emission filter 520; top optic; 30 flashes per well; settling time 0.5 s). The gain is set at 80–90% of the highest standard (1 000 ng/mL), and the blank is subtracted in the evaluation.
- In the reader software, fit a linear regression of the standards (S1–S3 and blank) (Calculation → Standard curve calculations → Linear regression fit); then, in Microplate View, select Linear Regression Fit under Recalculate concentration.
6.3 After finishing
- Clean the bench with 70% ethanol.
- Empty the small bench bin into the right waste bin.
- Wash your hands.
- Turn off equipment and lights.
- Lock doors, fridge and freezer.
7 Calculations
Sample concentration
\[c_\mathrm{sample} = \frac{c_\mathrm{well} \times 50}{1000}\]
- \(c_\mathrm{sample}\): concentration of the sample (ng/µL)
- \(c_\mathrm{well}\): concentration read from the standard curve (ng/mL)
- 50: dilution of the sample in the well (2 µL in 100 µL; dimensionless)
- 1000: conversion from ng/mL to ng/µL
Worked example (illustrative value): a well reads 400 ng/mL → 400 × 50 / 1000 = 20 ng/µL.
8 Quality control
| Control | Purpose | Acceptance criterion |
|---|---|---|
| Standard curve (S1–S3 + blank, in duplicate) | Calibration | R² ≥ 0.99; if not, prepare the standards again and repeat the curve before reading the samples |
| Samples within the curve | Valid reading | Above S1 (1 000 ng/mL in the well): dilute in 1X TE and measure again. Below the lowest standard: report as below the curve |
| DNA concentration of extracts | Enough DNA for PCR | General rule: ≥ 10 ng/µL success; < 5 ng/µL fail; depends on the starting quantity and parent material; between 5 and 10 ng/µL the PCR result decides |
9 Data and metadata
- Outputs: dsDNA concentration (ng/µL) per well; plate-reader export with the standard curve.
- Where: the Sample_Tracking sheet of the Metabarcoding Lab Pipeline workbook (DNA_ngul) or the project’s records, and the lab book; instrument use in the QC user log.
10 Safety and waste
Follow the manufacturer’s instructions and safety data sheets for each reagent. Biological samples are potentially hazardous and are handled following the facility’s procedures. EnGeL lab rules:
Before you enter: wear your lab coat, use appropriate gloves, closed shoes only, tie back long hair.
Not allowed: eating or drinking; touching phones or laptops with contaminated gloves; bags or coats on benches.
Label all reagents (name, concentration, date) and treat all chemicals as hazardous.
PicoGreen is dissolved in DMSO, which passes through skin, and binds DNA: wear gloves; keep away from light [TO CONFIRM: hazards from the SDS].
Waste
EnGeL waste streams (choose the right bin):
- Biological solid waste (black bin, white bag): contaminated gloves, contaminated disposable materials, solid biological laboratory waste, contaminated PPE. No liquids or sharps.
- Specific hazardous waste (red bin, red bag): chemicals and pharmaceuticals designated for this waste stream, specific hazardous laboratory waste, anatomical material; sharps only in a rigid sharps container. No loose sharps.
- Biological liquid waste (jerrican, UN 3291): liquid biological waste and contaminated biological liquids only. Use a funnel; rinse the original container and the funnel with dH2O and discard the rinse water in the same jerrican.
For this protocol: plates, tubes and tips go to biological solid waste (black bin) [TO CONFIRM].
11 Troubleshooting
| Problem | Likely cause | Solution |
|---|---|---|
| Standard curve R² < 0.99 | Pipetting error in the standards, or old lambda working stock | Prepare the standards again and repeat the curve. |
| Low signal overall | Reagent exposed to light, or glass tubes used | Prepare the working reagent fresh, in the dark, in plastic tubes. |
| Samples above S1 | Concentration beyond the curve | Dilute the samples in 1X TE and measure again; correct for the dilution. |
12 Remarks
The manual uses a 200 µL assay (100 µL sample + 100 µL working reagent); the lab uses half volumes (100 µL), with the same high-range standards (1 µg/mL, 100 ng/mL, 10 ng/mL, blank). For very dilute samples the manual gives a low-range curve (25 ng/mL to 250 pg/mL).
13 References
Thermo Fisher Scientific (Invitrogen) (n.d.). Quant-iT PicoGreen dsDNA Reagent and Kits, user guide, Pub. No. MAN0001931, Rev. A.0.
14 Version history and citation
| Version | Date | Change | By |
|---|---|---|---|
| 1.0.0 | 2026-11-20 | First version as a separate protocol, split from EnGeL-MOLB-NAQC 1.3.0 (PicoGreen part, from Andreia Quaresma’s “Quality Control: PicoGreen dsDNA quantification (96-well plate)” v1.0, 2026, checked against the Quant-iT PicoGreen user guide) at the request of the lab technicians. | Luís Cunha |
| 1.1.0 | 2026-11-20 | Samples above the standard curve diluted and measured again (confirmed by Luís Cunha); Luísa Fraga: testing and adjustments. | Luís Cunha |
| 1.1.1 | 2026-11-20 | Action when the standard curve fails: repeat the curve (confirmed by Luís Cunha). | Luís Cunha |
How to cite
Quaresma A, Fraga L, Cunha L (2026). Double-stranded DNA concentration in 96-well plates with PicoGreen (FLUOstar Omega plate reader). EnGeL-UC protocol EnGeL-MOLB-PICOGREEN, version 1.1.1. Zenodo. DOI [TO CONFIRM].
Licence
Text: CC BY 4.0. Code: MIT. Logos are the property of their institutions and are not covered by the licence.
15 Open questions
- Kit catalogue number; hazards from the SDS; waste route.
- DOI.