Double-stranded DNA concentration in 96-well plates with PicoGreen (FLUOstar Omega plate reader) · version 1.1.1 (2026-11-20). Full protocol: EnGeL-MOLB-PICOGREEN. Tick each step when done; write deviations in Notes.
| Date |
|
| Operator |
|
| Plate ID / samples |
|
| PicoGreen kit lot |
|
PicoGreen standards.
| S1 |
50 |
0 |
50 |
1000 |
| S2 |
5 |
45 |
50 |
100 |
| S3 |
0.5 |
49.5 |
50 |
10 |
| Blank |
0 |
50 |
50 |
0 |
Before starting
| ☐ |
• |
Use filter tips at all steps; plastic tubes only. |
| ☐ |
• |
Keep the PicoGreen reagent and working reagent protected from light. |
Plate and reading
| ☐ |
1 |
Pipette each standard into the black plate in duplicate (table in Reagent preparation). |
| ☐ |
2 |
Samples: 2 µL sample + 48 µL 1X TE per well + 50 µL working reagent. |
| ☐ |
3 |
Mix by pipetting or a short plate shake. |
| ☐ |
4 |
Incubate for 2 min at room temperature, protected from light. |
| ☐ |
5 |
Plate reader: Manage Protocol → “PicoGREEN_LC_100”; Start measurement → Gain Adjustment; then Start measurement. The reader protocol is a fluorescence-intensity endpoint read (Greiner 96 F-bottom plate; excitation filter 485-12, emission filter 520; top optic; 30 flashes per well; settling time 0.5 s). The gain is set at 80–90% of the highest standard (1 000 ng/mL), and the blank is subtracted in the evaluation. |
| ☐ |
6 |
In the reader software, fit a linear regression of the standards (S1–S3 and blank) (Calculation → Standard curve calculations → Linear regression fit); then, in Microplate View, select Linear Regression Fit under Recalculate concentration. |
After finishing
| ☐ |
• |
Clean the bench with 70% ethanol. |
| ☐ |
• |
Empty the small bench bin into the right waste bin. |
| ☐ |
• |
Wash your hands. |
| ☐ |
• |
Turn off equipment and lights. |
| ☐ |
• |
Lock doors, fridge and freezer. |
Notes / deviations