Double-stranded DNA concentration with the Qubit 4 fluorometer (dsDNA HS and BR assays) · version 1.2.0 (2026-11-20). Full protocol: EnGeL-MOLB-QUBIT. Tick each step when done; write deviations in Notes.
| Date |
|
| Operator |
|
| Plate ID / samples |
|
| Qubit dsDNA HS or BR kit lot |
|
Before starting
| ☐ |
• |
Use filter tips. |
| ☐ |
• |
Use only Qubit assay tubes, and label them on the lid, never on the side. |
Measurement
| ☐ |
1 |
Prepare the two standards in Qubit assay tubes: 190 µL working solution + 10 µL DNA standard #1; 190 µL working solution + 10 µL DNA standard #2. |
| ☐ |
2 |
For each sample, prepare a Qubit assay tube with 199 µL working solution + 1 µL sample. |
| ☐ |
3 |
Vortex vigorously for 3–5 s without creating bubbles. |
| ☐ |
4 |
Incubate all tubes at room temperature for 2 min. |
| ☐ |
5 |
On the Home screen, touch dsDNA, select dsDNA High Sensitivity (or dsDNA Broad Range with the BR kit) and touch Read standards. |
| ☐ |
6 |
Insert Standard #1, close the lid and touch Read standard; when the reading is complete (c. 3 s), remove it. Repeat with Standard #2. |
| ☐ |
7 |
Touch Run samples, select the sample volume (1 µL) and the output units. |
| ☐ |
8 |
Insert each sample tube, close the lid and select Read tube; remove the tube when the reading is complete (c. 3 s). Repeat for all samples. |
After finishing
| ☐ |
• |
Clean the bench with 70% ethanol. |
| ☐ |
• |
Empty the small bench bin into the right waste bin. |
| ☐ |
• |
Wash your hands. |
| ☐ |
• |
Turn off equipment and lights. |
| ☐ |
• |
Lock doors, fridge and freezer. |
Notes / deviations