Illumina library quantification by qPCR (KAPA Library Quantification Kit)

Authors

Andreia Quaresma

Luísa Fraga

Luís Cunha

Published

2026-11-20

Protocol ID EnGeL-MOLB-KAPA
Version 1.2.0
Date 2026-11-20
Status draft
Area Molecular biology (MOLB)
Authors Andreia Quaresma (lead; prepared the protocol; ORCID 0000-0002-8678-5800); Luísa Fraga (ORCID 0000-0003-3153-4593); Luís Cunha (review and approval; ORCID 0000-0002-5870-2537)
Reviewed and approved by Luís Cunha
Laboratory Environmental Genomics Laboratory (EnGeL-UC), Environmental Ecology & Ecotoxicology Group (E3G), Centre for Functional Ecology, Department of Life Sciences, University of Coimbra · Laboratório Associado TERRA
Contact luis.cunha@uc.pt · ewormuc@gmail.com
Licence Text: CC BY 4.0 · Code: MIT
Main hazards GHS08 Health hazard KAPA Library Quantification Kit: may cause damage to organs (H371; kit safety data sheet)
Downloads Protocol: PDF · Word. Bench sheet: PDF · web page

1 Purpose and scope

This protocol determines the molar concentration of Illumina sequencing libraries (pooled amplicon libraries) by qPCR, before dilution for sequencing. It is used at the end of library preparation (EnGeL-MOLB-MB). It measures only molecules carrying the Illumina P5/P7 adapter sequences.

2 Principle

The library is diluted into the range of six DNA standards of known concentration. Library and standards are amplified with primers that bind the P5 and P7 adapter sequences, using a SYBR Green qPCR master mix. The Ct values of the standards give a standard curve, from which the concentration of each library dilution is read and corrected for the difference in size between the standard (452 bp) and the library (KAPA Library Quantification Kit, Roche).

3 Before you start

Item Details
Hands-on time 1 h
Total time 2 h 30 min
Sample 1 µL of the pooled, bead-cleaned library (for the 1:1 000 intermediate dilution); average fragment size measured on the TapeStation; PicoGreen or Qubit concentration known
Prepare in advance Combined Master Mix + Primer Mix (first use of the kit); plate layout; standards thawed on ice
Key safety points See Safety and waste

4 Materials

4.1 Equipment

  • qPCR instrument: QuantStudio 3 (Applied Biosystems)
  • Pipettes and filter tips
  • Vortex mixer and mini-centrifuge

4.2 Consumables

  • qPCR plate and optical seal
  • Low-binding tubes

4.3 Reagents and kits

  • KAPA Library Quantification Kit for Illumina (Roche) [TO CONFIRM: catalogue number], containing:
    • KAPA SYBR FAST qPCR Master Mix (2X)
    • Primer Mix (10X)
    • DNA Standards 1–6
  • Library dilution buffer: 10 mM Tris-HCl pH 8.0–8.5 + 0.05% Tween 20 (called “TE pH = 8” in the source) [TO CONFIRM: name and recipe]
  • PCR-grade water
  • Positive control, 200 nM [TO CONFIRM: product and catalogue number]

5 Reagent preparation

  • Combined Master Mix + Primer Mix: see the first steps of the Procedure (first use of the kit only).
  • Library dilution buffer: 10 mM Tris-HCl pH 8.0–8.5 + 0.05% Tween 20 [TO CONFIRM: preparation, storage and shelf life].

6 Procedure

6.1 Before starting

  • Clean the working space and material with bleach and water and leave the UV light on for at least 15 min.
  • Use filter tips at all steps.
  • Clean forceps and spatulas with 3% sodium hypochlorite and 96% ethanol between samples.

6.2 Quantification

Kit preparation (first use)

  1. Add the 1 mL Primer Mix (10X) to the 5 mL bottle of SYBR FAST Master Mix (2X); vortex briefly.
  2. Store the combined mix at −20 °C (or 4 °C for short-term use, per kit insert); keep away from light.
  3. Thaw the standards on ice, vortex and spin down. Do not dilute the standards further.

Library dilutions (serial, in library dilution buffer)

Dilution Make from Volume taken (µL) Dilution buffer (µL) Use
1:1 000 Pool 1 999 Intermediate only
1:10 000 1:1 000 100 900 Run in qPCR
1:20 000 1:1 000 50 950 Run in qPCR
1:100 000 (optional) 1:10 000 10 90 If PicoGreen > c. 50 nM
  1. Prepare the dilutions above. Vortex and spin each dilution before taking the next step. Check against the PicoGreen value: the diluted library should be about 0.2–20 pM, inside the standard range.

Positive control: treat the positive control like a library. It starts at 200 nM and is diluted in the same series: 1:1 000 (intermediate), then 1:10 000 and 1:20 000, which should give about 20 pM and 10 pM in the reaction.

Reaction set-up (20 µL per well)

Component Per well (µL)
Master Mix + Primer Mix (combined) 12
PCR-grade water 4
Diluted library, standard or NTC buffer 4
  1. Prepare a master mix of 12 µL combined mix + 4 µL water per well, plus 10% excess.
  2. Dispense 16 µL per well, then add 4 µL of diluted library, standard or dilution buffer (for the NTC).
  3. Seal, spin briefly, and keep the plate in the dark until loading.

Plate layout (96-well): Standards 1–6 in triplicate (18 wells); NTC (dilution buffer as template) in duplicate (2 wells); positive control and each library at 1:10 000 and 1:20 000 in triplicate (6 wells each). With the positive control, up to 11 libraries fit on one plate (18 + 2 + 6 + 6 × 11 = 92 wells). In the software, enter the standards as Standard with quantities 20, 2, 0.2, 0.02, 0.002 and 0.0002 pM (S1–S6).

Cycling

Step Temperature Time Cycles
Initial denaturation 95 °C 5 min 1
Denaturation 95 °C 30 s 35
Annealing/extension + read 60 °C 45 s
Melt curve 65 °C, then to 95 °C at 0.15 °C/s 15 s at each temperature 1

Ramp rate 1.6 °C/s for all other steps (instrument settings in use).

  1. Run the plate on the qPCR instrument (QuantStudio 3) with the cycling above.
  2. Export the results as .csv (tab Export) and check the run against the run QC criteria before calculating.

6.3 After finishing

  • Clean the bench with 70% ethanol.
  • Empty the small bench bin into the right waste bin.
  • Wash your hands.
  • Turn off equipment and lights.
  • Lock doors, fridge and freezer.

7 Calculations

On the Results sheet of the exported file, use the Quantity Mean (pM) reported for each dilution:

\[c_\mathrm{lib} = \frac{Q \times \dfrac{452}{L} \times D}{1000}\]

  • \(c_\mathrm{lib}\): library concentration (nM)
  • \(Q\): Quantity Mean reported by the software for that dilution (pM)
  • 452: size of the kit DNA standard (bp)
  • \(L\): average fragment size of the library (bp), from the TapeStation
  • \(D\): dilution factor (e.g. 20 000; dimensionless)
  • 1000: conversion from pM to nM

Calculate \(c_\mathrm{lib}\) for each valid dilution, then average them.

Worked example (illustrative values): Quantity Mean 1.5 pM at 1:20 000, library size 595 bp: 1.5 × (452 / 595) × 20 000 / 1000 = 22.8 nM.

8 Quality control

Accept the run before calculating:

Check Accept
Standard curve efficiency 90–110% (slope −3.6 to −3.1)
Standard curve R² ≥ 0.99
ΔCt between consecutive standards c. 3.3 (3.1–3.6)
Replicate Ct SD ≤ 0.2–0.3; omit a clear outlier
NTC Undetermined
ΔCt between 1:10 000 and 1:20 000 c. 1.0 (0.8–1.2)
Sample Ct Within the Ct range of Standards 1–6
Melt curve Single peak, no primer-dimer peak
Positive control (200 nM) Measured within ±20% of 200 nM (160–240 nM)
Expected results
Melt curve plot with derivative fluorescence against temperature; library and control curves share one peak, the standards peak at a slightly higher temperature, and the no-template control is flat.
Figure 1: Melt curve of a good run: the libraries (here three pools at 1:10 000 and 1:20 000, labelled “1:10” and “1:20” in the software) and the positive control give one peak each, the DNA standards a single peak at a slightly higher temperature, and the NTC stays flat. Data: EnGeL-UC (Andreia Quaresma).
Standard curve: Ct against log quantity for six standards, falling on a straight line from Ct 9 to Ct 26.
Figure 2: Standard curve of a good run (Standards 1–6 in triplicate): slope −3.505, R² 0.999, efficiency 92.9%, within the acceptance criteria above. Data: EnGeL-UC (Andreia Quaresma).

9 Data and metadata

  • Outputs: exported qPCR results (.csv); library concentration (nM) per library, averaged over valid dilutions.
  • Where: the Library_QC sheet of the Metabarcoding Lab Pipeline workbook [TO CONFIRM] and the lab book; instrument use in the QuantStudio 3 user log.
Minimum to record
  • Library / pool ID and EnGeL label
  • Run date, operator, instrument (QuantStudio 3)
  • KAPA kit lot number
  • Dilutions run; Quantity Mean per dilution; library size used (bp); final concentration (nM)
  • Run QC values (efficiency, R², NTC)

10 Safety and waste

Follow the manufacturer’s instructions and safety data sheets for each reagent. Biological samples are potentially hazardous and are handled following the facility’s procedures. EnGeL lab rules:

  • Before you enter: wear your lab coat, use appropriate gloves, closed shoes only, tie back long hair.
  • Not allowed: eating or drinking; touching phones or laptops with contaminated gloves; bags or coats on benches.
  • Label all reagents (name, concentration, date) and treat all chemicals as hazardous.

Waste

EnGeL waste streams (choose the right bin):

  • Biological solid waste (black bin, white bag): contaminated gloves, contaminated disposable materials, solid biological laboratory waste, contaminated PPE. No liquids or sharps.
  • Specific hazardous waste (red bin, red bag): chemicals and pharmaceuticals designated for this waste stream, specific hazardous laboratory waste, anatomical material; sharps only in a rigid sharps container. No loose sharps.
  • Biological liquid waste (jerrican, UN 3291): liquid biological waste and contaminated biological liquids only. Use a funnel; rinse the original container and the funnel with dH2O and discard the rinse water in the same jerrican.

For this protocol: used qPCR plates, tips and tubes go to biological solid waste (black bin) [TO CONFIRM].

11 Troubleshooting

Problem Likely cause Solution
Diluted library outside the standard range Wrong dilution for the library concentration Use the PicoGreen value to choose dilutions; add the 1:100 000 dilution if PicoGreen > c. 50 nM.
ΔCt between dilutions not c. 1.0 Pipetting or dilution error Repeat the dilutions, vortexing and spinning each one.
Extra melt-curve peak Primer dimers or non-specific product [TO CONFIRM]; see also the kit technical data sheet (Roche, KR0405).
Signal in the NTC Contamination [TO CONFIRM]

12 Remarks

The standard is 452 bp; the size correction uses the average fragment size measured on the TapeStation.

13 References

Roche (n.d.). KAPA Library Quantification Kit – Illumina Platforms. Technical data sheet KR0405, v8.17. Troubleshooting: https://rochesequencingstore.com/wp-content/uploads/2017/10/KAPA-Library-Quantification-Kit-for-Illumina.pdf

14 Version history and citation

Version Date Change By
1.0.0 2026-11-16 First EnGeL-UC version, from the lab protocol “KAPA qPCR library quantification” (2026). Luís Cunha
1.0.1 2026-11-17 Contributors: Luísa Fraga and Andreia Quaresma. Luís Cunha
1.1.0 2026-11-17 Updated from Andreia Quaresma’s v1.0 (2026-11-05): times, sample volume, positive control treated like a library (200 nM, 1:1 000 then 1:10 000 and 1:20 000), diluted library range 0.2–20 pM, melt-curve and ramp settings, plate capacity, hazard H371 (GHS08), technical data sheet KR0405 v8.17, expected-results figures (melt curve, standard curve). Luís Cunha
1.1.1 2026-11-17 Authorship: all who contributed are authors, with their roles; lead author first where identified, reviewer (Luís Cunha) last. Luís Cunha
1.2.0 2026-11-20 Positive-control acceptance criterion confirmed by Luís Cunha: within ±20% of 200 nM (160–240 nM). Luís Cunha

How to cite

Quaresma A, Fraga L, Cunha L (2026). Illumina library quantification by qPCR (KAPA Library Quantification Kit). EnGeL-UC protocol EnGeL-MOLB-KAPA, version 1.2.0. Zenodo. DOI [TO CONFIRM].

Licence

Text: CC BY 4.0. Code: MIT. Logos are the property of their institutions and are not covered by the licence.

15 Open questions

  1. Authorship (author or contributor model for the handbook, pending); DOI.
  2. Kit catalogue number.
  3. Positive control: product and catalogue number.
  4. Library dilution buffer: the source calls it “TE pH = 8” but gives 10 mM Tris-HCl + 0.05% Tween 20 (no EDTA). Name, preparation and storage?
  5. Where results are recorded (Lab Pipeline Library_QC sheet?).
  6. Troubleshooting: action for extra melt-curve peaks and for signal in the NTC.
  7. Waste route for used plates.