Bench sheet · EnGeL-MOLB-KAPA

Illumina library quantification by qPCR (KAPA Library Quantification Kit) · version 1.2.0 (2026-11-20). Full protocol: EnGeL-MOLB-KAPA. Tick each step when done; write deviations in Notes.

Date
Operator
Plate ID / samples
KAPA Library Quantification Kit lot

Before starting

☐ No. Step
☐ • Clean the working space and material with bleach and water and leave the UV light on for at least 15 min.
☐ • Use filter tips at all steps.
☐ • Clean forceps and spatulas with 3% sodium hypochlorite and 96% ethanol between samples.

Quantification

☐ No. Step
☐ 1 Add the 1 mL Primer Mix (10X) to the 5 mL bottle of SYBR FAST Master Mix (2X); vortex briefly.
☐ 2 Store the combined mix at −20 °C (or 4 °C for short-term use, per kit insert); keep away from light.
☐ 3 Thaw the standards on ice, vortex and spin down. Do not dilute the standards further.
Dilution Make from Volume taken (µL) Dilution buffer (µL) Use
1:1 000 Pool 1 999 Intermediate only
1:10 000 1:1 000 100 900 Run in qPCR
1:20 000 1:1 000 50 950 Run in qPCR
1:100 000 (optional) 1:10 000 10 90 If PicoGreen > c. 50 nM
☐ No. Step
☐ 4 Prepare the dilutions above. Vortex and spin each dilution before taking the next step. Check against the PicoGreen value: the diluted library should be about 0.2–20 pM, inside the standard range.
Component Per well (µL)
Master Mix + Primer Mix (combined) 12
PCR-grade water 4
Diluted library, standard or NTC buffer 4
☐ No. Step
☐ 5 Prepare a master mix of 12 µL combined mix + 4 µL water per well, plus 10% excess.
☐ 6 Dispense 16 µL per well, then add 4 µL of diluted library, standard or dilution buffer (for the NTC).
☐ 7 Seal, spin briefly, and keep the plate in the dark until loading.
Step Temperature Time Cycles
Initial denaturation 95 °C 5 min 1
Denaturation 95 °C 30 s 35
Annealing/extension + read 60 °C 45 s
Melt curve 65 °C, then to 95 °C at 0.15 °C/s 15 s at each temperature 1
☐ No. Step
☐ 8 Run the plate on the qPCR instrument (QuantStudio 3) with the cycling above.
☐ 9 Export the results as .csv (tab Export) and check the run against the run QC criteria before calculating.

After finishing

☐ No. Step
☐ • Clean the bench with 70% ethanol.
☐ • Empty the small bench bin into the right waste bin.
☐ • Wash your hands.
☐ • Turn off equipment and lights.
☐ • Lock doors, fridge and freezer.