Illumina library quantification by qPCR (KAPA Library Quantification Kit) · version 1.2.0 (2026-11-20). Full protocol: EnGeL-MOLB-KAPA. Tick each step when done; write deviations in Notes.
| Date |
|
| Operator |
|
| Plate ID / samples |
|
| KAPA Library Quantification Kit lot |
|
Before starting
| ☐ |
• |
Clean the working space and material with bleach and water and leave the UV light on for at least 15 min. |
| ☐ |
• |
Use filter tips at all steps. |
| ☐ |
• |
Clean forceps and spatulas with 3% sodium hypochlorite and 96% ethanol between samples. |
Quantification
| ☐ |
1 |
Add the 1 mL Primer Mix (10X) to the 5 mL bottle of SYBR FAST Master Mix (2X); vortex briefly. |
| ☐ |
2 |
Store the combined mix at −20 °C (or 4 °C for short-term use, per kit insert); keep away from light. |
| ☐ |
3 |
Thaw the standards on ice, vortex and spin down. Do not dilute the standards further. |
| 1:1 000 |
Pool |
1 |
999 |
Intermediate only |
| 1:10 000 |
1:1 000 |
100 |
900 |
Run in qPCR |
| 1:20 000 |
1:1 000 |
50 |
950 |
Run in qPCR |
| 1:100 000 (optional) |
1:10 000 |
10 |
90 |
If PicoGreen > c. 50 nM |
| ☐ |
4 |
Prepare the dilutions above. Vortex and spin each dilution before taking the next step. Check against the PicoGreen value: the diluted library should be about 0.2–20 pM, inside the standard range. |
| Master Mix + Primer Mix (combined) |
12 |
| PCR-grade water |
4 |
| Diluted library, standard or NTC buffer |
4 |
| ☐ |
5 |
Prepare a master mix of 12 µL combined mix + 4 µL water per well, plus 10% excess. |
| ☐ |
6 |
Dispense 16 µL per well, then add 4 µL of diluted library, standard or dilution buffer (for the NTC). |
| ☐ |
7 |
Seal, spin briefly, and keep the plate in the dark until loading. |
| Initial denaturation |
95 °C |
5 min |
1 |
| Denaturation |
95 °C |
30 s |
35 |
| Annealing/extension + read |
60 °C |
45 s |
|
| Melt curve |
65 °C, then to 95 °C at 0.15 °C/s |
15 s at each temperature |
1 |
| ☐ |
8 |
Run the plate on the qPCR instrument (QuantStudio 3) with the cycling above. |
| ☐ |
9 |
Export the results as .csv (tab Export) and check the run against the run QC criteria before calculating. |
After finishing
| ☐ |
• |
Clean the bench with 70% ethanol. |
| ☐ |
• |
Empty the small bench bin into the right waste bin. |
| ☐ |
• |
Wash your hands. |
| ☐ |
• |
Turn off equipment and lights. |
| ☐ |
• |
Lock doors, fridge and freezer. |
Notes / deviations